Moudi Cascade Impactor Model 100 &110 Features & Applications
Precision size classification of aerosols for
- air pollution
- pharmaceutical aerosol drug delivery
- diesel blow by and engine emission testing
- automotive airbag testing
- industrial hygiene studies
Sharp cut-point characteristics
Four equal geometrical intervals per decade of particle size
Models 100 and 110
- 30-L/min sampling flow rate
- 8 and 10 impactor stages + after filter
- 0.18µm and 0.056µm minimum cut-point
Model 100 & 110 Cut-Point Performance 
Model 115 Moudi Cut-Point Performance 
*The Model 115, 3-stage Nano-MOUDI™ is available in a non-rotating version only. The Model 115 and Model 125 requires a separate 6 CM/H vacuum pump (NAN-090) to operate.