Universal Air Sampler
Model 310 UAS™ - Request a Quote
The Model 310 is a 300 L/min, medium flow air sampler consisting
of an inlet, a PM10 and a PM2.5 virtual impactor for size separation
of airborne particles, a final filter, and a polyurethane foam
(PUF) sampler to collect organic vapors downstream of the filter.
The modular design allows the system to be used for PM10, PM2.5, or PM10/PM2.5 dichotomous sampling. The system also includes a blower, blower speed control, and instrumentation for flow measurement.
A seven-day solid-state timer allows up to 14 programmable starts and stops of the system. For instance, the timer can be programmed to collect day-time samples from 6:00 a.m. to 6:00 p.m. each day for a whole week, so that a weekly total sample can be obtained.
FEATURES & APPLICATIONS
- Air pollution and air quality research
- PM10, PM 2.5, and PM10/PM2.5 dichotomous sampling
- 300-L/min sampling flow rate
- Virtual impactor size classification
- Solid-state motor speed control for flow adjustment
- 7 day solid state timer allowing up to 14 programmable starts
and stops per week
SPECIFICATIONS
(subject to change without notice)| Sampling flow rate |
300 liters per minute |
| Filter size |
8" x 10" (major flow), 3" x 8" (minor
flow) |
| Virtual impactor cutpoints |
2.5 and 10 µm standard (1.0 µm optional) |
| Dimensions (L x W x H) |
762 x 1524 x 1422 mm (30" x 60" x 56") |
| Weight | 46 kg (96 lb) |
| Power |
115 VAC, 60 Hz, 10A (230 VAC, 50 Hz, 5A optional) |