Atomic Absorption Spectrophotometer BK-AA860B Pro

BK-AA860B Pro features an advanced 8-lamp design, a flame atomizer, 7-level spectral bandwidth adjustment, and Zeeman background correction. It delivers full-wavelength, full-element analysis with high-precision background compensation, capable of detecting over 30 elements at concentrations ranging from mg/L(10⁻⁶) to μg/L(10⁻⁹) levels. Ideal for geochemistry, metallurgy, materials science, agriculture, petrochemicals, third-party testing, and academic research.

$0.00

SKU AP-SPC-AA-860B-PRO Category Tag

Description

Features:

1. The flame system uses high-precision mass flow controllers for precise automated flow control.

2. The constant transverse magnetic field Zeeman background correction technology provides superior full-spectrum background absorption correction capability.

3. Built-in cooling water detection and temperature monitoring ensure reliable operation of the magnetic field system.

4. The patented quick-release burner head design solves cleaning challenges, allowing tool-free disassembly in just 3 seconds.

5. A complete gas-liquid separation system and fuel filtration unit ensure stable operation in high-humidity environments or with limited gas supply.

6. Multiple safety interlock mechanisms with automatic shutdown and alarm functions protect against operational anomalies.

7. The modular instrument design and circuit architecture enable easy maintenance and quick component replacement.

Technical Parameters:

Model BK-AA860B Pro
Wavelength Range 170~900nm
Spectral Bandwidth 0.1, 0.2, 0.4, 0.8, 1.2, 1.6, 2.6nmnm
Wavelength Accuracy ±0.1nm
Wavelength Repeatability ≤0.05nm
Detector Photomultiplier tube detector
Monochromator Classic Czerny-Turner configuration
Optical System Focal Length 430mm
Baseline Stability Drift ≤0.0005Abs, noise ≤0.0005Abs(Steady 60min)

Drift ≤0.001Abs, noise ≤0.001Abs(Dynamic 30min)

Resolution Spectral bandwidth deviation ≤0.02nm

Manganese double line valley to peak energy ratio ≤25%

Characteristic Concentration Flame method for Cu: ≤0.03μg/mL/1%
Detection Limit Flame method for Cu: ≤0.002μg/mL
Preciseness Flame method for Cu: ≤0.25%
Background Correction Ability Zeeman correction: ≥1.8Abs, background ≥150 times
Diffraction Grating Plane diffraction grating(1800 grooves/mm, blazed at 250nm)
Lamp Stand 8(Support simultaneous preheating of 1~4 lights)
Standard Accessory Element lamp(Cu*1, Hg*1), software working station
Optional Accessory Wide slit combustion head, adjustable nebulizer, green energy-saving accessories, audit trailing software, oil-free air compressor, computer, printer, recirculating chiller, flame autosampler
Power Supply AC220V, 50Hz
External Size(W*D*H) 1010*620*630mm
Package Size(W*D*H) 1220*800*910mm
Net Weight 115kg
Gross Weight 190kg

Flame System:

Air-acetylene Burner 100mm all-titanium burner
Atomization Chamber System Integrally molded from polyphenylene sulfide
Magnetic Field Strength ≥1.0T
Reading Mode Foot pedal operation
Safety System Emergency shutdown protection switch, automatic gas cut-off/shutdown/alarm in case of abnormal status, abnormal air pressure, ignition failure, gas leakage, water seal, abnormal flow rate, etc.

Software Working Station:

Smart Optimization One-key operation for instrument optimization, supporting multi-task analysis
Repeat Measurement 1~99 replicates with auto-calculation of mean/SD/RSD
Print/Export Reports Word/Excel supported
Auto-Calibration Curve fitting, re-slope&concentration calculation, incl. standard addition method