Features 1. Compatible with both normal-phase and reversed-phase solvents, suitable for use in GPC systems, and supports solvent exchange. 2. An optimized, high-efficiency heat dissipation structure ensures stable system operation under demanding condition. 3. The fully upgraded sealing design minimizes system performance degradation and consistently delivers accurate test results. 4. The front panel is coated with a corrosion-resistant finish, enhancing the product’s durability against a wide range of solvents. 5. Compatible with Kromstation and Rubikstation chromatography workstations (network versions), compliant with data integrity regulations. D1100+ UV-visible Detector
| Item | Specification |
| Wavelength Range | 190~700nm |
| Light Resource | Deuterium lamp |
| Display | 2*16 LCD |
| Spectral Bandwidth | 8nm |
| Wavelength Accuracy | ±1nm |
| Wavelength Repeatability | ≤1nm |
| Set Fan of Wavelength-time Program | 0.1~999.9min |
| Response Time | 0.1~5.0s |
| Noise | ≤±1×10⁻⁵ AU (Empty flow cell, 254nm, constant ambient temperature) |
| Drift | ≤1.5×10⁻⁴ AU/h (Empty flow cell, 254nm, constant ambient temperature) |
| Linearity Range | ≥1.8AU (5%, 254nm) |
| Optical Path of Flow Cell | 10mm |
| Withstand Voltage of Flow Cell | ≤8MPa |
| Communication Mode | RS232 or USB |
P1100+ High-Pressure Constant Flow Pump
| Item | Specification |
| Flow Rate | 0.001 - 10.000 mL/min (Step: 0.001 mL/min) |
| Accuracy | ≤ ±0.3% (1.0 mL/min, 8 ± 2 MPa, water, ambient temperature) |
| Precision | ≤ 0.085% (1.0 mL/min, 8 ± 2 MPa, water, ambient temperature) |
| Max Pressure | 45MPa(0.001~5.000mL/min) 20MPa(5.001~10.000mL/min) |
| Pressure Accuracy | ±3% or <±0.8 MPa |
| Pressure Pulsation | ≤1% |
| Leak-proofing Capability | Pressure drop≤1.5MPa (40MPa, 10min) |
| Communication mode | RS-485 |
| Dimension/Net Weight | 420*260*160mm/14kg |
| Power Supply | AC220V, 50Hz |
| Power | 80W |
O1100+ Series Column Oven
| Item | Specification |
| Temperature Range | Ambient temperature+5~99°C |
| Setting Precision | 0.1°C |
| Temperature Accuracy | ≤±0.5°C |
| Temperature Stability | ≤0.2°C |
S1100+ Autosampler
| No. | Items | Specifications |
| 1 | Injection range | 0~100μL, 500μL, 2mL |
| 2 | Injection repeatability | RSD<0.3% Full sample loop injection (standard conditions) RSD<0.5% Partial volume injection (lossy injection, standard conditions) RSD<1.0% Partial volume injection (without loss, standard conditions) |
| 3 | Injection linearity | R²>0.999 |
| 4 | Sample residue | <0.01% |
| 5 | Highest pressure | 45MPa |
| 6 | Communication method | UDP network communication |
| 7 | Injection trigger | Digital mode, analog mode |
Technical Parameters:
| Model | Agress1100+(Type I) | Agress1100+(Type II) |
| Pump | P1100+*1 | P1100+*2 |
| Detector | D1100+*1 | |
| Sampler Injector | Rheodyne 7725i*1 | |
| Valve Mounting Frame | VB1100+*1 | |
| Solvent Bottle Tray | T1100+*1 | |
| Workstation | Kromstation*1 | |
| System Toolkit | Agress1100+*1 | |
| Column | Supersil ODS2 5μm 4.6*250mm*1 | |
| Gradient Mixer | / | TD-1-15*1 |
| Power Supply | AC220V±10%, 50Hz | |
| Optional Accessory | Autosampler, column, column oven, evaporative light scattering detector, differential refractive index detector, fluorescence detector, solvent bottle | |
| Packing Size(W*D*H) | 580*400*280mm 580*400*280mm 580*400*280mm | 580*400*280mm 580*400*280mm 580*400*280mm 580*400*280mm |
| Gross Weight | 16kg 16kg 8kg | 16kg 16kg 16kg 8kg |
Product relationship
Available Through Applied Physics
Applied Physics supplies and supports quotation of this product. Product branding and manufacturer information, where stated on the product page or supporting documentation, remain authoritative.
Applied Physics portfolio: Medical
Catalog relationship: Available Through Applied Physics
Related Applied Physics resources
Continue through the Medical portfolio
Use these technical and portfolio pages to compare applications and related equipment. To keep this exact product attached to an RFQ, use the product’s existing Request a Quote or Add To Order control.