Scanning Uv Vis Visible Spectrophotometer

The UV-VIS Spectrophotometer is a high-precision analytical instrument designed for quantitative and qualitative analysis in laboratories, research institutions, environmental monitoring, and chemical testing.

$3,995.00

SKU AP-BK-SPV-4 Category Tag

Description

Features:

1. With SiO2 coating optical mirror, reducing the pollution from outside fully.

2. Adopted high-class grating with wholly hermetic light path design, to ensure the instrument has the super low stray light.

3. With GLP self-check function, check the wavelength accuracy and photometry accuracy.

4. Real-time monitoring the lifetime of Deuterium lamp and Tungsten lamp with advanced system.

5. Wavelength calibration, wavelength setting, change lamp source and dark current calibration automatically.

Basic Functions:

1. Photometry

Test Abs., transmittance and energy by fixed wavelength.

2. Quantitative

Linear fit and linear fit through zero two modes.

a. Coefficient, standard sample input and standard Sample read three modes to establish standard curve.

b. Establish A=K1*C+K0, can search original data, graph curve, parameters settings.

c. Can save 240 group curves, can test 240 data in each curve.

d. Double wavelength, Triple wavelength test functions.

3. Kinetics

Used for time course scanning or reaction rate calculations△A/t, can search all data.

4. Multi-Wavelength

Can test Transmittance and Abs. with 8 different wavelengths at most.

5. Scanning

User can set the scan range from 190nm to 1100nm to test the max. Abs. peak value, can do derivation, arithmetical operationsto the graph.

6. Biology

6 methods: DNA/Protein, UV, Lowry, BCA,CBB and Biuret.

Technical Parameters:

Model BK-UV1900

BK-UV1900PC

BK-V1900

BK-V1900PC

Optical System Single Beam, Grating 1200 lines/mm
Wavelength Range 190~1100nm 320~1100nm
Spectral Bandwidth 2nm
Wavelength Accuracy ±0.5nm
Wavelength Repeatability ≤0.2nm
Photometric Accuracy ±0.002A(0~0.5Abs), ±0.004A(0.5~1.0Abs), ±0.5%T(0~100%T)
Photometric Repeatability 0.001Abs(0~0.5Abs), 0.002Abs(0.5~1.0Abs), ≤0.2%T(0~100%T)
Stray Light ≤ 0.04%T@360nm; 220nm ≤ 0.04%T@360nm
Stability ± 0.001A/h@500nm
Baseline Flatness ± 0.002A
Noise 0.0005Abs@500nm
Display 480*272 65 thousand true color TFT LCD
Photometric Mode T, A, C, E
Photometric Range 0~200%T, -0.301~3.0A
Detector Silicon Photodiode
Light Source Deuterium Lamp, Tungsten Lamp
Input Membrane Keypad
Output USB Print and data output, Connect PC
Standard Accessories 1cm glass cuvette*4, 1cm quartz cuvette*2, 1cm quadruple cuvette holder*1, Scanning software(only for PC type) 1cm glass cuvette*4, 1cm quadruple cuvette holder*1, Scanning software(only for PC type)
Optional Accessories 1cm standard single slot cuvette holder, long optical path cuvette holder, cuvette, micro cuvette, test tube type absorber rack, solid sample holder, deuterium lamp(only for UV series),tungsten lamp, quadruple cuvette holder, printer, four-slot water bath type constant temperature absorption pool, X-Y adjustable micro absorption pool, single bath type constant temperature absorption pool, program-controlled thermostatic absorption pool seat, 8 racks, reflection accessories, general application software or scanning software(not for PC type)
Power Supply AC85~265V, 50/60Hz
Package Size(W*D*H) 620*500*370mm
Gross Weight 18kg 16.5kg

Additional information

Model

BK-UV1900, BK-UV1900PC, BK-V1900, BK-V1900PC

Reviews

There are no reviews yet.

Only logged in customers who have purchased this product may leave a review.