Description
Features: 1. Automatic 6 position cell holder for 10-50mm optical path. 2. The base adopts a thickened and reinforced design, improving the stability of the optical path and ensuring measurement accuracy. 3. Sealed design to avoid external influences on the optical path. Extend the service life of the instrument. 4. Long life deuterium lamp,tungsten lamp (no less than 2000 hours), socket design, easier to replace. 5. 1200 line/mm holographic gratings, moisture-proof coated optical mirrors, improving the measurement accuracy and service life of the product. 6. Supports USB data output and printing, supports Bluetooth. 7. Provide various specialized accessories to expand the application range of the instrument. 8. Base function: photometric, quantitative, kinetic,multi wavelength. 9. PC software design meets GMP requirements, multi user management, audit tracking function (Optional). Technical Parameters:
| Model | BK-UV6T |
| Optical System | 1200 line/mm Grating, Single Beam |
| Wavelength Range | 190~1100mm |
| Spectral Bandwidth | 2.0nm |
| Wavelength Accuracy | ±0.5nm(@D656.1±0.1nm) |
| Wavelength Repeatability | ≤0.2nm |
| Photometric Accuracy | ±0.002A(0-0.5Abs), ±0.004A(0.5-1.0Abs), ±0.2% T(0-100% T) |
| Photometric Repeatability | ±0.001A(0-0.5Abs), ±0.002A(0.5-1.0Abs), ±0.1%T(0-100%T) |
| Stray Light | ≤0.04T@360nm; 220nm |
| Stability | ±0.001A/h@500nm |
| Baseline Flatness | ±0.001A |
| Noise | ±0.001Abs |
| Display | 1024*600 7 inch touch screen |
| Photometric Mode | T, A, C, E |
| Photometric Range | 0-200%T,-4.0-4.0A,0-9999.9C |
| Detector | Silicon Photodiode |
| Light Source | Deuterium Lamp, Tungsten Lamp |
| Input | Touch Screen |
| Interface | USB-A*2, USB-B*1, Blue Tooth |
Product relationship
Available Through Applied Physics
Applied Physics supplies and supports quotation of this product. Product branding and manufacturer information, where stated on the product page or supporting documentation, remain authoritative.
Applied Physics portfolio: Medical
Catalog relationship: Available Through Applied Physics
Related Applied Physics resources
Continue through the Medical portfolio
Use these technical and portfolio pages to compare applications and related equipment. To keep this exact product attached to an RFQ, use the product’s existing Request a Quote or Add To Order control.





Reviews
There are no reviews yet.