Description
Benchtop pH/ORP/Ion Meter
Features:
pH:
1. 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers
2. Automatic electrode diagnosis shows the pH slope and offset
3. Solution temperature coefficient compensates for the pure water samples and references the pH to 25°C
ORP:
1. 1 point offset calibration allows adjusting the displayed value to a known standard
2. Relative and absolute millivolt modes ensure the reliable ORP measurements
Ion Concentration:
1. 2 to 5 points calibration, including the 8 concentration points can be selected
2. Electrode management is capable of storing and recalling up to 3 electrode slopes
3. Automatic electrode diagnosis shows the calibration points and slopes
4. Selectable ion measurement methods (direct reading, known addition, known subtraction, sample addition, sample subtraction) and concentration units (ppm, mg/L, mol/L or mmol/L)
Water Hardness:
1. 2 to 5 points calibration from low to high concentrations
2. Selectable measurement units (German degree, English degree, French degree, mmol/L and mg/L) are used for professional water hardness measurements
General Features:
1. Automatic temperature compensation ensures accurate readings over the entire range
2. Auto-Read function senses and locks the measurement endpoint
3. Timed Interval Readings sends the measurement data to a PC or printer
4. Limit Alarm automatically alerts when reading exceeds the specified range
5. Calibration due alarm prompts user to calibrate the meter regularly
6. Password protection prevents the unauthorized calibration and settings
7. Expanded memory stores or recalls up to 1000 data sets
8. Reset function automatically resumes all settings back to the factory defaults
Technical Parameters:
Model
PH-A120-CN
PH-A120-UK
PH-A130-CN
PH-A130-UK
PH-A131
PH
Range
-2.000~20.000pH
•
•
—
Resolution
0.001, 0.01, selectable
•
•
—
Accuracy
±0.002pH
•
•
—
Calibration Points
1 to 5 points
•
•
—
pH Buffer Options
USA, NIST, DIN, 5 custom buffers
•
•
—
ORP(optional)
mV Range
±2000.0mV
•
•
•
Relative mV Range
±2000.0mV
•
•
—
Resolution
0.1mV
•
•
•
Accuracy
±0.2mV
•
•
•
Calibration Points
1 point
•
•
—
Ion
Range
0.001~30000 (depending on the range of ISE)
—
•
•
Resolution
0.001, 0.01, 0.1, 1
—
•
•
Accuracy
±0.5% F.S. (monovalent), ±1% F.S. (divalent)
—
•
•
Measurement Units
ppm, mg/L, mol/L, mmol/L
—
•
•
Calibration Points
2 to 5 points (0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000)
—
•
•
Measurement Methods
Direct reading, known addition, known subtraction, sample addition, sample subtraction
—
•
•
Electrode Management
1 to 3 electrodes
—
•
•
Water Hardness
Range
0.05~200mmol/L, 0~1122°dH, 0~1404°e, 0~2000°fH, 0~8000mg/L (Ca2+)
—
•
•
Resolution
0.001, 0.01, 0.1, 1
—
•
•
Accuracy
±1% F.S.
—
•
•
Calibration Points
2 to 5 points (0.01, 0.1, 1, 10, 100mmol/L)
—
•
•
Temperature
Range
0~105°C/32~221°F
•
•
•
Resolution
0.1°C/0.1°F
•
•
•
Accuracy
±0.5°C/±0.9°F
•
•
•
Offset Calibration
1 point, reading ±10°C
•
•
•
Other Specifications
Temperature Compensation
0~100°C, manual or automatic
•
•
•
Solution Temperature Coefficient
25°C
•
•
—
Stability Criteria
Standard or high-accuracy
•
•
•
Calibration Due Alarm
1 to 31 days or off
•
•
•
Interval Readings
10, 30, 60 seconds, 10, 30 minutes or off
•
•
•
Password Protection
4 digits
•
•
•
Memory
1000 data sets
•
•
•
Communication Interface
USB
•
•
•
Connector
BNC, 3.5 mm jack socket
•
•
•
Display
7 inches TFT LCD
•
•
•
Power Supply
Standard: 220V 50/60Hz; Optional: 110V, 50/60Hz; DC12V power adapter
•
•
•
Outer Dimensions
240*220*80mm
Net Weight
1.7kg
Packing Size
320*250*200mm
Gross Weight
1.8kg
Remarks:
• means with; — means without.
Difference of PH-A120/130-CN& PH-A120/130-UK:
1. PH-A120/130-CN electrode is E201-BNC plastic body pH electrode, PH-A120/130-UK electrode is P11 glass pH electrode
2. PH-A120/130-UK is more accurate than PH-A120/130-CN
Ordering Information:
1. PH-A120/130-CN: Meter, E201-BNC plastic body pH electrode, temperature probe, pH buffer sachets, electrode holder and power adapter
2. PH-A120/130-UK: Meter, P11 glass pH electrode, temperature probe, pH buffer sachets, electrode holder and power adapter
3. PH-A131: Meter, ion selective electrode, temperature probe, standard solutions (100/1000ppm), ionic strength adjuster, electrode holder and power adapter