Description
Portable Conductivity/TDS/Salinity Meter PH-520 PH-530 PH-531 PH-540
PH-520 Features:
1. 1 to 3 points calibration with automatic recognition for conductivity standards.
2. Automatic electrode diagnosis shows the calibration points and factors .
3. Selectable cell constant, reference temperature, linear/non-linear and pure water compensation modes.
4. Automatic temperature compensation ensures accurate readings over the entire range.
5. Auto-Read function senses and locks the measurement endpoint.
6. Setup menu allows setting the number of calibration points, temperature unit, auto-power off, etc.
7. Reset function automatically resumes all settings back to the factory defaults.
8. Expanded memory stores or recalls up to 100 data sets.
9. USB communication interface for data transfer orconnecting a power adapter to meter.
PH-530/531/540 Features:
1. 1 to 5 points calibration with automatic recognition for conductivity standards.
2. Automatic electrode diagnosis shows the calibration points and factors.
3. Selectable cell constant, reference temperature, TDS factor, linear and pure water compensations, seawater and practical salinity measurement modes.
4. Automatic temperature compensation ensures accurate readings over the entire range.
5. Auto-Read function senses and locks the measurement endpoint.
6. Calibration due alarm prompts user to calibrate the meter regularly.
7. Auto-Power Off effectively conserves battery life.
8. Setup menu allows setting the number of calibration points, stabilitycriteria,temperature unit, date and time, etc.
9. Reset function automatically resumes all settings back to the factory defaults.
10. Expanded memory stores or recalls up to 500 data sets.
11. USB communication interface is easy to transfer data to PC.
12. Multi-mode powerscheme(batteries, power adapter and USB port) ensures that using the meter smoothly.
Accessories:
Meter, CON-1 conductivity electrode, temperature probe, standard solutions and carrying case.
Technical Parameters:
Model
PH-520
PH-530
PH-531
PH-540
Conductivity
Range
Standard
CON-1: 2μs/cm~20000μs/cm
Optional
CON-0.01: 0.01μs/cm~0.1μs/cm
CON-0.1: 0.1μs/cm~2μs/cm
CON-10: 20000μs/cm~200ms/cm
Accuracy
±0.5% F.S.
Resolution
0.001, 0.01, 0.1, 1
Calibration Points
1 to 3 points(PH-520), 1 to 5 points(PH- 530/531/540)
Calibration Solutions
10μS/cm, 84pS/cm, 1413pS/cm, 12.88mS/cm, 111.8mS/cm
TDS
Range
/
0~10.00, 100.0,
1000ppm, 10.00,
200.0ppt
/
0~10.00, 100.0,
1000ppm, 10.00,
200.0ppt
Accuracy
/
±1%F.S.
/
±1%F.S.
Resolution
/
0.01, 0.1, 1
/
0.01, 0.1, 1
TDS Factor
/
0.1~1.0(default 0.5)
/
0.1~1.0(default 0.5)
Salinity
Range
/
0.00~42.00psu, 0.00~80.00ppt
Accuracy
/
±1% F.S.
Resolution
/
0.01
Resistivity
Range
/
0.00~20.00MΩ
Accuracy
/
±1% F.S.
Resolution
/
0.01
Temperature
Range
0~105°C/32~221°F
Accuracy
±0.5°C/±0.9°F
Resolution
0.1°C/0.1°F
Offset Calibration
1 point, reading ±10°C
Other
Specifications
Temperature Compensation
0~100°C, manual or automatic
Temperature Coefficient
Linear(0.0~10.0%/°C), non-linear, pure water
Reference Temperature
20/25°C
Cell Constant
K=0.1, 1, 10
Stability Criteria
/
Low or high
Calibration Due Alarm
/
1 to 31 days or off
Hold Function
Manual or auto-endpoint
Auto-Off
30 minutes after last key pressed(PH-520), 10, 20 or 30 minutes after last key pressed(PH-530/531/540)
Memory
100 data sets(PH-520), 500 data sets(PH-530/531/540)
Communication Interface
USB
Connector
6-pin mini-DIN, 3.5mm jack socket
Display
Custom LCD(80*60mm)
Power Requirements
3*1.5V AA batteries or DC5V power adapter
Battery Life
Approximately 150 hours(Turn off the backlight)
Dimensions(mm)
170*85*30
Weight(kg)
300g
Package Size(W*D*H)mm
475*155*350
Gross Weight(kg)
2kg