Description
Vertical Pulse Vacuum Autoclave 50L 75L 100L
Features:
1. Steam water internal circulation.
2. Pulse vacuum exhaust to ensure the effect of steam sterilization.
3. Mobile probe detects to ensure the internal temperature of liquid directly, to ensure the sterilization effect of liquid program.
4. LCD screen display shows pressure, temperature, time and error codes.
5. Selectable programs for packed, unpacked, dressings, rubber, liquid, liquid custom, solid custom, sterilization of medium, dissolution of medium, B&D testing and vacuum testing etc.
6. Complete process of preheating, replacement, pulse exhaust, heating, sterilizing and drying run automatically.
7. 0.22μm Sterilization high efficiency air filter, ensuring that the vacuum-depleted air is sterile air to prevent re-pollution of sterile materials.
8. Standard built-in printer for recording the information during the entire sterilization process.
Technical Parameters:
Model
BKQ-B50V
BKQ-B75V
BKQ-B100V
Capacity
50L
75L
100L
Designed Pressure
-0.1/0.28MPa
Working pressure
0.23MPa
Designed Temp.
150℃
Working Temp.
105~134℃
Temp. Precision
0.1℃
Chamber Material
S30408
Chamber Size
φ386*490mm
φ386*670mm
φ386*900mm
Basket
φ360*260mm, 1pc
φ360*360mm, 1pc
φ360*330mm, 2pc
Vacuum System
Eject vacuum system
Noise
≤65dB(A)
Altitude
≤2000m, could be customized at 2000~3000m
Standard Accessories
Printer, mesh bottom basket
Consumption
5.5KW
6.2KW
Power Supply
AC220V, 50/60Hz(Standard); AC110V, 50/60Hz(Optional)
External Size(W*D*H)
700*610*1070mm
700*610*1300mm
Net Weight
127kg
130kg
150kg
Package Size(W*D*H)
820*740*1380mm
820*740*1530mm
Gross Weight
142kg
145kg
170kg






Reviews