Description
Veterinary Auto Hematology Analyzer BH-HA310VET
Application:
The Hematology Analyzer can detect abnormalities of healthy animals, assess sick animals and monitor sick animals undergoing treatment by performing routine blood tests. It is suitable for animal experiment center, drug research center, animal and plant quarantine department, zoo, pet hospital, animal husbandry and veterinary system and other institutions.
Features:
1. Detection many animal species, with upgrade function.
2. Seven animal detection modes(cow, horse, sheep, goat, rabbit, dog, cat, camel).
3. Automatic detection, eliminating manual error.
4. Abnormal results intuitive prompt, timely and accurate screening out abnormal experimental results.
5. Three detection modes: venous blood, peripheral blood, pre-diluted peripheral blood.
6. 9. 7-inch color touch screen, data status at a glance.
7. One button for maintenance.
8. It can be turned on continuously for 24 hours.
9. Applying cyanide-free hemolytic agent to sample, safe and environmental protection without stimulation.
10. Automatic system diagnosis and maintenance, with a variety of alarm functions.
11. Fully enclosed reagent line, no contamination.
12. Excellent data management and communication function, automatically store 100,000 results.
13. Supporting online remote diagnosis and quality control.
The BH-HA310VET hematology analyzer has three detection modes of venous blood, capillary blood, prediluted blood, including 21 detection parameters:
MID# GRAN# LYM% RBC MCV MCHC RDW-CV MPV PCT P-LCC WBC
LYM# MID% GRAN% HGB HCT MCH RDW-SD PLT PDW P-LCR
Technical Parameters:
Model
BH-HA310VET
Throughput
60 samples/hour
Principle
Impedance for WBC differentiation, WBC/RBC/PLT count, Colorimetric method for HGB
Channel
2
Parameters
21 parameters(including MID#, GRAN#, LYM%, RBC, MCV, MCHC, RDW-CV, MPV, PCT, P-LCC, WBC, LYM#, MID%, GRAN%, HGB, HCT, MCH, RDW-SD, PLT, PDW, P-LCR)
Histograms
3 histograms(including WBC, RBC, PLT)
Sample Volume
Venous blood: 10μl; Capillary blood: 10μl; Prediluted: 20μl
Calibration
Manual calibration, automatic calibration, and fresh blood calibration
Counting Mode
Venous blood, capillary blood, prediluted blood
Printing Model
Auto print, manual print
LIS System
Supporting Bi-directional LIS
Language
Chinese, English, Spanish, French
QC
L-J, X-B
Reagent
Diluent 10L, Hemolytic agent 250ml(animal specific), Probe cleaner 50ml
Date Input
9.7-inch touch screen; mouse, keyboard(optional)
Output
Built-in printer, supporting external printers
Printer Paper
57*30mm
Interface
4 USB ports, 1 network port, and 1 RS232 serial port
Storage
Automatic storage, 100,000 results with histograms(including sample information)
Information Transmission
Bidirectional LIS communication, HL7 and many other transmission protocols
Unclog
Reverse high-pressure flush
Work Temp.
10~40°C
Power Supply
AC100~240V, 50/60Hz
Relative Humidity
≤80%
Atmospheric Pressure
86.0kPa~106.0kPa
Carry-Over Rate
Parameter
WBC
RBC
HGB
PLT
CV
≤1.5%
≤1.0%
≤1.0%
≤3.0%
Background
WBC
RBC
HGB
PLT
≤0.5*109/L
≤0.05*1012/L
≤2g/L
≤10*109/L
Linearity
Parameter
Measurement Range
CV
WBC
(1~10.00)*109/L
(10.1~99.9)*109
±0.5*109/L
±5%
RBC
(0.3~1.00)*1012/L
(1.01~7.00)*1012/L
±0.5*1012/L
±5%
HGB
(20~70)g/L
(71~200.0)g/L
±2.0g/L
±3%
PLT
(20~100)*109/L
(101~999)*109/L
±10.0*109/L
±10%
Precision
Parameter
Repeatability
Measurement Range
WBC
≤ 4.0%
(3.5~9.5)*109/L
RBC
≤ 2.0%
(3.8~5.8)*1012/L
HGB
≤ 2.0%
(115~175)g/L
PLT
≤ 8.0%
(125~350)*109/L
HCT
≤3.0%
35%~50%
External Size(W*D*H)
295*505*430mm
Package Size(W*D*H)
Instrument: 404*564*616mm;
Reagent 1: 340*340*340mm;
Reagent 2: 200*150*100mm
Net Weight
20kg
Gross Weight
Instrument: 23kg; Reagent 1: 13kg; Reagent 2: 13kg











Reviews